EDXtreme User Instructions
59
Radio communications not
working at all
Dead batteries. Distance
is excessive, dead radio
pocket
Bring remote closer to dynamometer. Allow several seconds to retrain.
Operating channels mis-
matched
Remote and link must be on the same operating channel. See EDXtreme
and Communicator configurations of COM1 for radio (under Comm menu)
and Common Configurations.
Excessive radio noise or
interference in
environment
Remove dynamometer and Communicator from the environment. Attempt
communications in an area free of local radio signals. See Radio
Information section of the manual.
Remote reading changes to
dashes
Low batteries, lost
communications
See steps above for improving communications.
Display locks up on
DILLON marquee
Poor connection
between EDX and
Communicator
Remove batteries from EDX and Communicator, replace them and power
up.
Problem
Possible Cause
Solution