EDXtreme User Instructions
41
The following screen is displayed:
6.
Select
Yes
if you only want the instrument to automatically shut down when
there is inactivity. Select
No
if you want the instrument to power down after the
timer counts down, regardless of any activity.
5.1.6 Config--ChPwd
The next soft key is
ChPwd
. Use this to change the password used to access the
configuration menu.
IMPORTANT: Changing the password denies access to this menu without entry
of the new password. Contact your Dillon distributor if you lose your password.
5.1.7 Config--Reset
The last soft key in this set is the
Reset
soft key. Use this to reset the system to its
factory default settings.
This concludes the Configuration menu section.
5.1.8 Config--Reso
Reso
stands for resolution. Resolution is the value by which the EDX displays
increments. Press this key to set the unit to standard (1000 divisions) resolution or
enhanced (5000 divisions) resolution. For example: 10,000 lbf would count by
10 lbf (10,000 lbf/1000 divisions = 10 lbf) in standard resolution
2 lbf (10,000 lbf/5000 divisions = 2 lbf) in enhanced resolution
Standard resolution may have 1000 or 1250 divisions. Enhanced resolution may have
4000 or 5000 divisions. This follows normal display increment practice of changing by
1, 2, 5 or a multiple or submultiple of those numbers (e.g. 10, 20, 50, .1, .2, .5, etc.)
Motion Resets Timer? (Yes)
Yes
No
Summary of Contents for EDXtreme
Page 10: ...8 EDXtreme User Instructions 1 8 Declaration of Conformity EDX and Communicator II ...
Page 11: ...EDXtreme User Instructions 9 1 9 Declaration of Conformity EDX PSU ...
Page 12: ...10 EDXtreme User Instructions ...
Page 34: ...32 EDXtreme User Instructions 4 1 6 Setpts Future Feature ...
Page 63: ...EDXtreme User Instructions 61 10 6 Good Force Measurement Practice ...
Page 67: ......