20
EDXtreme User Instructions
Press the
Mode
soft key and the display changes to peak measurement mode. This
mode shows the peak force applied to the EDX since the last peak clearing action.
Delete the peak reading by pressing the
Clear
soft key.
Press the
Mode
soft key again and the display shows live force and peak readings
simultaneously.
Press the
Mode
soft key again and the display returns to the force measurement mode.
3.2 Force Measurement
Follow these steps to perform a gross force measurement.
1.
Turn on the unit with the
On/Off
key.
2.
Remove any weight from the EDX.
3.
Zero the EDX by pressing the
ZERO
key.
4.
Apply the force to the EDX and read the gross force on the display.
You can change the units of measure of the display by pressing the
Units
soft key. See
the note below. Zero reference is maintained after the instrument is powered off and
will be recalled with the next power-on. Zero reference may be lost if battery power is
removed.
3.2.1 Force Measurement Rezero and Net Force Display
Rezeroing allows the weight or load of fixturing to be invisible to the measurement. The
zeroed load must always be considered as part of the maximum capacity.
1.
Turn on the unit with the
On/Off
key.
2.
Remove any weight from the EDX.
3.
Zero the EDX by pressing the
ZERO
key.
4.
Apply the tare force to the EDX and press the
ZERO
key.
5.
Apply the force to the EDX and read the net force on the display.
Steps
2
and
3
are not required if the weight of the fixturing is not needed. This should
be maintained, however, to know cumulative loads.
Power up display modes may be configured. See Config>Mode section.
Unit of measure can be changed only if multiple units are enabled in the Configuration
menu.
Summary of Contents for EDXtreme
Page 10: ...8 EDXtreme User Instructions 1 8 Declaration of Conformity EDX and Communicator II ...
Page 11: ...EDXtreme User Instructions 9 1 9 Declaration of Conformity EDX PSU ...
Page 12: ...10 EDXtreme User Instructions ...
Page 34: ...32 EDXtreme User Instructions 4 1 6 Setpts Future Feature ...
Page 63: ...EDXtreme User Instructions 61 10 6 Good Force Measurement Practice ...
Page 67: ......