4.4 Absolute and Linearity Ranges of Measured Parameters
Within linearity range, the instrument is guaranteed to provide the specified accuracy.
Beyond this linearity range, the instrument is able to display results, but may not
guarantee accuracy defined in specifications.
If the value is over the maximum range of guaranteed linearity, the instrument cannot
measure it and the result will be marked with an E (Error) flag.
To measure a sample, whose parameters exceed the maximum value indicated in
the table below, pre-dilution is recommended. See section 5.3.1.3 of this manual.
The linearity ranges of primary parameters in normal measuring mode:
Parameter Linearity
Ranges
Maximum
Unit
WBC
0...75 100
10
9
cells/liter
RBC
0...15 20
10
12
cells/liter
PLT
0...700 1000
10
9
cells/liter
HGB
0...250 400 g/l
HCT
0...100 - %
MCV
30...150 - fl
MPV
3...30 - fl
Table 2. Linearity ranges of parameters