Abacus junior 5
User’s Manual
19
2.3.3 Preparing for shipment
The second item in Exit (Shutdown) menu should be used when the instrument is to
be shipped or left unused for a longer time. The instrument will ask you to utilize the
cleaning tube kit and 100ml of distilled water.
Follow the instructions appearing on the display.
EXIT
PREPARING FOR SHIPMENT (2)
Here, the user is instructed to remove tubing
connectors, so they could be drained.
Leave the waste connector attached.
Next, you should connect the cleaning tube
kit to the reagent inputs, submerging the free
end in a bottle containing at least 100 ml of
distilled water.
The analyzer will flush any remaining
reagents from the system into the waste
container.
As a next step, the analyzer asks you to
remove the cleaning tube kit.
When finished, the analyzer prompts you to
power off the system. Remove the waste
connector after shutting down.
2.3.4 Emergency handling
In case of emergency situation - e.g. smoke or fire is experienced (short circuit), the -
cut off power immediately and use a fire-extinguisher.
! Message 5001/19300
Remove reagent tubing at rear reagent
inputs (Diluent, Lyse, Cleaner, EOS, BAS)
! Message 5002/19300
Connect min 100 ml distilled water to
reagent inputs using cleaning tube kit.
! Message 5003/19300
Remove cleaning tube kit. Keep reagent
inputs free.