14
© Diatron MI Ltd. 2005
2 INSTALLATION
2.1 General information
This chapter provides instructions for the installation of the
Abacus junior 5
hematology analyzer. The procedures described below must be followed correctly to
ensure proper operation and service. Please carefully read and follow all instructions
in this
User’s Manual
before attempting to operate
Abacus junior 5.
Abacus junior 5
hematology analyzer is a precision instrument, and must be
handled accordingly. Dropping or other improper handling of the instrument will
disturb calibrated mechanic and electronic components and/or cause other damage.
Always handle the instrument with care.
2.2 Environmental factors
Locate
Abacus junior 5
so that it will not be exposed to extreme temperature
variations. Temperature should be held relatively constant to obtain maximum
reliability. The ambient temperature range for the instrument is 15°C to 35°C.
Do not place the instrument close to any open windows, ovens, hot places, radiators,
air conditioners, or in direct sunlight. Do not expose the instrument to vibrations from
equipment such as centrifuges and shaker baths or alike.
Abacus junior 5
external power supply must be plugged into a
grounded AC outlet.
Do not plug the instrument into electrical outlets on the same
circuit as devices that operate intermittently, and use large electrical currents such as
air conditioners, refrigerators, compressors, etc. It is not advisable to use extension
cords, especially multiple outlet extensions. If you are located in an area that
experiences excessive power fluctuations or are using a generator, connect the
instrument to a surge protector (filtered surge protector preferred).
Abacus junior 5
should be installed on a flat, level surface with adequate
room for the reagents, the optional keyboard and printer. The backside of the
instrument should be at least 6 inches from adjacent walls or other equipment to
allow sufficient airflow and ventilation.
Abacus junior 5
uses external reagent containers. Care should be taken so
that the table holding the instrument is not pushed against the wall blocking the tubes
between the instrument and reagent containers.