Appendix A
Error Messages
D:
Difference error: The difference between parallel
measurements is greater than the permissible maximum
difference.
C:
Curve error: There is a jump in the optical curve.
T:
Out of range error: If the measurement time is outside the
measuring range (Min time – Max time).
R:
Calibration data error: There is no calibration data in the
calibration menu.
I:
Incubation is too long: If the measurement does not start in
time because the incubation time is too long.
O:
Incubation was overheated: If the thermostatic block is too warm.
E:
Expired lot: The reagent used has expired.
R:
Barcode type error: If the reagent type has changed
(eg. Dia-PT ->Dia-PT LIQUID), but the control range still
belongs to the old reagent type.
Q:
Out of QC: The control result is out of control range.
S:
Slope error: Slope is valid during D-Dimer measurement, but
the user chose not to re-measure the sample in dilution. In
this case ΔOD is not displayed since µgFEU/ml >5, and S is
listed as an error.
d
Diluted sample: D-Dimer sample was measured in dilution.