DLPT2+
Operating Instructions
30
3.6
Changing the Insulation Test Voltage
The Insulation test performed in the CLI and CLII
test sequences can be changed to 250V for
appliances with MOVs which would fail an Insulation
test performed at 500V.
In order to change the Insulation voltage press the
SETUP button to enter setup mode. Once in setup
mode press the CLII button, the selected insulation
voltage will be displayed next to the
Riso
enunciator.
To leave setup mode press the SETUP button.
3.7
Changing the CLI Leakage limit
The CLI Leakage pass/fail limit can be changed
between 0.75mA and 3.5mA.
In order to change the Leakage limit press the
SETUP button to enter setup mode. Once in setup
mode press the LKGE button, the selected pass/fail
limit will be displayed next to the
Ileak
enunciator.
To leave setup mode press the SETUP button.