DLPT2+
Operating Instructions
25
3.3
Testing a Class II Appliance
•
Visually inspect the appliance and mains cord
for signs of damage.
•
If the appliance passes a visual inspection
proceed with the electrical tests.
•
Plug the earth test lead into the Earth Continuity
Test Terminal on the DLPT2+ end panel.
•
Plug the appliance into the DLPT2+ front panel
test/mains outlet socket.
•
Connect the earth test probe to an exposed
metal part on the appliance.
•
If the Appliance under test has an ON/OFF
switch, make sure it is in the ON position.
•
Press the CLII test key
•
The display will momentarily show the insulation
test voltage before proceeding with the test.
Note: The power switch on the appliance under
test must be in the ON position to perform
an insulation test. If no appliance is
detected the DLPT2+ will display the
following.