MFC-CB™ Operation and Maintenance Manual
©2001~2007
DH Instruments, a Fluke Company
Page 72
VIN(
n
)
Purpose
Read the MFC measurement voltage if in voltage mode.
Syntax “VIN
n
”
Arguments
n
:
Device channel number (optional). Uses the active device
channel if not given.
'1' to address DEV1
'2' to address DEV2
Remarks
The MFC-CB can measure the voltage signal from an MFC/MFM. The
MFC-CB must be in voltage mode before this can be done (see the
"MFCCH" command). The returned data is always in volts.
Example Command:
Reply:
“VIN2
“1.0137 V"
Errors ERR#
10:
ERR# 43:
The prefix "
n"
is invalid
Not in volts mode
See Also
3.2.2, “DEV”
VOUT(
n
)(=)
Purpose
Read or set the voltage set point if in voltage mode.
Syntax “VOUT
n
”
“VOUT
n
=volts”
Arguments
n
:
volts:
Device number (optional). Uses the active device if not
given.
'1' to address DEV1
'2' to address DEV2
The voltage to be output (0 to 6 volts).
Defaults “VOUT
n
=0”
Remarks
The MFC-CB can output a voltage setpoint. The initial voltage is set, and
then corrected if needed according to the measurement of the "sense" input.
The MFC-CB must be in voltage mode before this can be done (see the
“MFCCH” command). The data is always in volts.
Example Command:
Reply:
“VOUT1=2”
“2.0000 V”
Errors ERR#
7:
ERR# 10:
ERR# 43:
The voltage is invalid
The prefix "
n"
is invalid
Not in volts mode
See Also
3.2.8, 7.1.1, "MFCCH", "VSENSE"