ANALOG OUTPUT SPECIFICATIONS
VOLTAGE OUTPUT
Range
0 to 6 VDC
Uncertainty
±
0.015 %FS
Resolution
0.1 mVDC
Output Impedance (max)
100
Ω
per channel
Load Impedance (min)
10 k
Ω
per channel
Short Circuit Protection
Yes (indefinite)
CURRENT OUTPUT
Range
4 to 20 mA
Uncertainty
±
0.025 %FS
Resolution
0.4
μ
A
Maximum Load
500
Ω
(compliance voltage: 10 VDC)
Short Circuit Protection
Yes (indefinite)
1.2.3
ANALOG INPUT SPECIFICATIONS
VOLTAGE INPUT (MEASURE AND SENSE)
Range
0 to 6 VDC
Min to Max Measurable
-0.25 to 6.00 VDC
Uncertainty
±
0.015 %FS
Resolution
0.1 mVDC
Input Impedance (min)
1 M
Ω
CURRENT INPUT (MEASURE)
Range
4 to 20 mA
Min to Max Measurable
4 to 20 mA
Uncertainty
±
0.025 %FS
Resolution
0.4
μ
A
Input Impedance
250
Ω
1.2.4
VALVE TEST POINT SPECIFICATIONS
Range
+2 to +15 VDC (in reference to -15 VDC)
Accuracy
±
0.1 %FS
Resolution
0.3 mVDC