3. OPERATION
Page
29
©2001~2007
DH Instruments, a Fluke Company
PRINCIPLE
MFC-CB set points are entered using
[MFC]
. In some cases, it can be useful to adjust or
trim the set point slightly, directly from the run screen without going through set point entry.
The MFC-CB jog function makes it easy to adjust the set point slightly. Pressing the
[
←
]
or
[
→
]
key when in the MAIN run screen causes the set point value to increase or
decrease by the amount specified in
[SETUP]
,
<2jog>
(see Section 3.3.2).
Use [SET MFC] to enter a new set point (see Section 3.2.8).
OPERATION
The MFC set point value can be jogged from the MAIN run screen or from the POWER screen.
To jog the set point, press
[
←
]
to decrease the value or
[
→
]
to increase the value.
Each time
[
←
]
or
[
→
]
is pressed, the set point value is adjusted up or down by the value
specified in
[SETUP]
,
<2jog>
(see Section 3.3.2).
Use
[SET MFC]
to enter a completely new set point (see Section 3.2.8).
See Section 3.3.2 for information on setting the jog increment values.
3.3
[SETUP]
PURPOSE
To access a menu of commonly used MFC-CB functions and features that DO NOT have direct
function keys.
PRINCIPLE
The
[SETUP]
key accesses a menu of commonly used MFC-CB functions and features that do
NOT have direct function keys. These functions include:
•
MFC
(see Section 3.3.1):
To create, store and edit the MFC profiles available for selection
from
[MFC]
.
•
jog
(see Section 3.3.2):
To adjust the set point jog increment value.
•
res
(see Section 3.3.3):
To adjust the resolution of MFC-CB set point and
measurement displays.
•
flowU
(see Section 3.3.4):
To customize the flow unit choices available when setting up
MFC profiles.
•
adj
(see Section 3.3.5):
To adjust the MFC set points and/or measurements by an
adder
and/or a
multiplier
.