OPERATION
To enter a set point for the active device channel, press
[SET MFC]
from the run screen.
The display is:
1. Active device channel indicator.
2. Unit of measure.
3. Entry field for set point value. Defaults to last
entry or zero.
Use the numerical and editing keys to enter the set point value desired. Press
[ENTER]
when the desired set point is displayed to return to the run screen with the new set
point active.
If the set point value entered results in an out of range electrical output value, MFC-CB
displays an error message. Press [ESCAPE] or [ENTER] to return to the <Enter set
point:> screen and enter a valid set point value. Consider that a set point that appears
valid may cause an out of range condition due to the application of a K factor or
adjustment value (see Sections 3.2.1, 3.3.5 and 7.1.1).
If the active MFC profile is not a controller, an error message appears when a set point
value is entered. To enter set point values, use MFC profile #1 (for V) or #2 (for mA) or
set up and select an MFC profile for a controller (see Section 3.3.1).
The set point value may be adjusted incrementally directly from the run screen using the
[
←
] or [
→
] keys to “jog” the set point (see Section 3.2.9).
To view a summary of the active MFC profile, press [+/-] from the run screen.
Set point entries are device channel specific. The set point entered on one device channel
does not affect the other device channel.
3.2.9
[
←
] AND [
→
] (JOG MFC SET POINT)
PURPOSE
To adjust (jog) the MFC set point by a small value directly from the run screen, rather
than by entering a new set point using
[SET MFC]
.
See Section 3.3.2 for information on setting the jog increment values.
2
Enter set point:
1.0000 V D1