MFC-CB™ Operation and Maintenance Manual
©2001~2007
DH Instruments, a Fluke Company
Page 36
3.4.1
<1RESET>
PURPOSE
To reset various MFC-CB settings to default or factory values.
PRINCIPLE
MFC-CB stores user definable settings in non-volatile memory. The reset menu allows
the user to selectively or completely reset these values to factory defaults. Resets clear
settings that the user may have made, and should be used only to restore the MFC-CB to
a known state. MFC-CB will go through its reboot routine after any type of reset
is executed.
OPERATION
To access the reset choices press
[SPECIAL]
and
select
<1reset>
. The display is:
Select the desired reset. After confirmation, the reset occurs. A reset always puts the
MFC-CB through its power up routine as if power had been turned OFF and back ON.
See Sections 3.4.1 through 3.4.5 for detailed information on the specific reset choices.
RESET functions change user settings that affect set point and read signals. If not used
properly, resetting can cause out of tolerance measurements. RESET functions should
only be used by qualified personnel with reference to this manual.
3.4.1.1
<1SETS>
PURPOSE/OPERATION
To access Reset - Sets, press
[SPECIAL]
,
<1reset>
,
<1sets>
.
Reset - Sets clears and sets to default the following user settings:
•
DISPLAY mode to CLEAN (see Section 3.2.4.6).
•
Device channel selection to DEV1 (see Section 3.1.2.4).
•
K factor OFF and value to 1.0000 (see Section 3.2.1).
•
Unit to Volts (V) (see Section 3.2.2).
•
DEV1 and DEV2 MFC profile selection to #1 (see Section 3.2.6).
•
All drivers (optional) OFF and drive operation to toggle (see Section 3.2.7).
•
Jog value to 0.1 %FS (see Section 3.3.2).
•
Resolution to 0.01 %FS (see Section 3.3.3).
1sets 2units 3MFC
4cal 5all