43
6.
After a large population of results has been acquired for a given doctor and IOL model,
navigate to the Personalized Lens Constants Table to view the Procedures that were
performed by the selected doctor using the selected lens. Review the Procedures that have
“
Use Record
”
selected and modify as necessary. See step 5 for considerations on filtering
which results to use.
7.
The
“
New Personalized Constants
”
group box shows the results of the Personalized Lens
Constants calculation based on the Procedures that are currently selected in the table. The
“
Current Personalized Constants
”
group box displays the lens constants that are currently
being used for the selected Doctor and IOL.
8.
Click the
[↑]
button to replace the
“
Current Personalized Constants
”
with the
“
New
Personalized Constants
”.
These new Personalized Lens Constants will now be available for
use in future IOL power calculations.
NOTE:
Automatically calculating lens constants is similar to the procedure described above
except
that the lens constants are recalculated as soon as
“
Add to Personalized Constants
Table
”
checkbox is selected.