![DGH TECHNOLOGY SCANMATE FLEX Operator'S Manual Download Page 11](http://html1.mh-extra.com/html/dgh-technology/scanmate-flex/scanmate-flex_operators-manual_2491685011.webp)
11
B
-
SCAN PROBES
1.
Probe Lens
2.
Probe Top Identifier
3.
Probe Handle
4.
Probe Connector
DGH1912 Label
DGH1920 Label
MODEL NUMBER
DGH1912
DGH1920
TRANSDUCER
12.5 MHz (nominal)
20 MHz (nominal)
SCAN ANGLE DISPLAYED
60 degrees
60 degrees
PROBE POSITION ACCURACY
±3 degrees
±3 degrees
FOCUS
20 mm ±2 mm
21 mm ±3 mm
DEPTH OF FIELD
15 mm—30 mm
15 mm —
30
mm
LATERAL RESOLUTION
0.5 mm ±0.1 mm
0.4 mm ±0.1 mm
AXIAL RESOLUTION
<0.35 µS at focal point
<0.13 µS at focal point
REPLACEMENT PROBE CABLE
FLEX
-
PCABLE
(8’
-
0”
length)
FLEX
-
PCABLE
(8’
-
0”
length)
NOTE:
Probe may be completely submerged in liquid when disconnected from probe cable.
1
2
3
4
A
-
SCAN PROBE—
DGH 6006
1.
Probe Lens w/ Fixation LED
2.
Immersion Shell Stop
3.
Probe Handle
4.
Cord Strain Relief
5.
Probe cord to cable
DGH 6006 Engraving
TRANSDUCER
10 MHz (nominal)
FOCAL POINT
LED brightness adjustable 0%
-
100% in software.
AXL MEASUREMENT RANGE
15.0 mm —
40.0
mm
ACD MEASUREMENT RANGE
2.0 mm —
6.0
mm
LT MEASUREMENT RANGE
2.0 mm —
7.5
mm
AXL MEASUREMENT REPEATABILITY
± 0.03 mm StDev (immersion) , ± 0.10 mm StDev (contact)
AXL MEASUREMENT ACCURACY
100 microns (immersion) , 180 microns (contact)
REPLACEMENT PROBE CABLE
N/A. Permanently attached.
(6’
-
0”
length)
NOTE:
Probe may be submerged up to the edge of the probe cord strain relief.
1
2
4
3
5