![DGH TECHNOLOGY SCANMATE FLEX Operator'S Manual Download Page 10](http://html1.mh-extra.com/html/dgh-technology/scanmate-flex/scanmate-flex_operators-manual_2491685010.webp)
10
USB INTERFACE MODULE —
FLEX
-
USBIM
1.
Battery Charge Check Button
2.
Battery Charge Indicator Lights (blue)
3.
Active Charging Indicator Light (amber)
4.
Power On Indicator Light (green)
5.
Battery Enable/ Disable Switch
6.
B
-
Scan / UBM connector
7.
A
-
Scan connector
8.
A
-
Scan test block
9.
USB Connector
10.
Power Switch
11.
AC/DC Transformer Connector
Bottom Panel Label
Front Panel
MAX FRAMES PER SECOND (FPS)
40 fps, 128 sectors, 2
-
way scan mode
20 fps, 128 vectors, 1
-
way scan mode
20 fps, 256 vectors, 2
-
way scan mode
12 fps, 256 vectors, 1
-
way scan mode
PIXEL COUNT
512 x 512 (4:1 oversampling)
SAMPLING FREQUENCY
80 MHz (max)
PC CONNECTION
USB 2.0
BATTERY PACK TYPE*
10.8 V, 3250 mAh Li
-
Ion
BATTERY LIFE
Approximately 4 hours when fully charged.
INPUT VOLTAGE
18 Vdc
INPUT CURRENT
1.67 A
POWER SUPPLY MODEL**
EMMAQ180167
-
P5P
-
IC (Medical Grade AC/DC Adapter)
* Battery only to be serviced/replaced by DGH Technology, Inc.
** Isolation from electrical supply mains provided by using the listed medical grade AC/DC adapter.
NOTE:
Do not submerge housing. Device housing not sealed against water ingress.
SYSTEM COMPONENTS
The Scanmate Flex is capable of operating various probes. The following tables describe the
Scanmate Flex device and supported probe specifications.
3
4
2
1
6
7
8
9
10
11
4
3
2
1
5