3300-0298-0M Rev E
y
Solace Dialysis Scale Owner’s
14
DIALYSIS OPERATION WITH PATIENT ID, CONT.
14.
Once the amount of fluid removed from the patient is equal to or greater than the TARGET
FLUID REMOVAL weight value, the background of the display will change to red, and an
alarm (a beeping sound) will turn on.
W
WEIGHT: P
POUNDS/GROSS
A
AC
5.0
15.
When the display changes to red and the alarm (beeping sound) turns on, the operator can
touch anywhere on the display to end the Dialysis process. The display will change to a
Results
screen with details of the Dialysis process shown.
R
Reesults
Date: 2/24/2020
Target: 5
Actual: 5.0
Exit
16.
With the
Results
screen displayed, the operator can press
, to print a ticket with the
details of the Dialysis process, or press
Exit
to return to the main screen.
17.
Assist the patient off the scale.