Solace Dialysis Scale Owner’s
19
Settings Page 3
SETTINGS
PAGE 3
Capacity 1: 600.0
Capacity 2: 1000.0
Interval 1: 2
Interval 2: 5
<<<
Exit
>>>
Capacity 1:
Touching this key will open the capacity screen. This will allow you to set the capacity of
the first range. Using the numeric keys, enter a new setting, and then press the
Save
key.
The first range of the scale should be set to 600.0.
Capacity 2:
Touching this key will open the capacity entry screen for the second weight range. This
setting MUST be greater than Capacity 1 if used. Using the numeric keys, enter a new
setting, and then press the
Save
key.
The second range of the scale should be set to 1000.0.
NOTE:
To disable the second weight range, set Capacity 2 to a setting of (0) zero.
Interval 1:
Touching this key will open the interval entry screen. This will allow you to set the scale
interval for the first range. Using the numeric keys, enter a new setting, and then press the
Save
key.
Allowable settings for Interval 1 are 1, 2, or 5.
Interval 2:
If a second range capacity has been set, then Interval 2 will be used as the interval of the
second weight range if enabled (see Capacity2). Using the numeric keys, enter a new
setting, and then press the
Save
key.
Allowable settings for Interval 2 are 1, 2, or 5.
x
Press
the
>>>
key to advance to Settings Page 4.
x
Press
the
<<<
key to return to Settings Page 2.
x
Press
the
Exit
key to exit Setup and return to the Weight screen.