3300-0298-0M Rev E
y
Solace Dialysis Scale Owner’s
4
AC POWER CONNECTION
The Solace In-Floor Dialysis Scale has been designed to operate from a connection to a
permanent power source and contains an internal power supply that can be easily connected
to
100 to 240 VAC 50/60Hz.
1.
Referring to the figure below, remove the conduit knockout from the enclosure, and
replace it with a conduit connector per local Electrical Code.
2.
Trim the jacket back 2-inches from the end of the power supply cable, and strip the
insulation back from the three individual wires approximately 3/8 inch.
3.
Run AC Hot, Neutral, and Ground wires through the conduit into the box area. Trim and
strip these as required to connect with the power supply cable.
4.
Using supplied wire nuts connect the 3-wires from the AC power source to the 3-wires
from the power supply.
5.
Remember that the Black wire connects to Brown, and a White wire connects to Blue.
IMPORTANT!
Permanently connected equipment
must have a readily accessible
disconnect device incorporated in the
fixed wiring, such as a circuit breaker.
Power Su
pp
ly
Power Supply Cable
Customer furnished wire
from power source
Green
Blue
Black
Green
White
Brown
Use supplied wire nuts
to connect power supply
cable to customer
furnished wire.
Conduit connector
furnished by customer