Solace Dialysis Scale Owner’s
27
To Enter Calibration, Cont.
SETTINGS
PAGE 5
Calibrate Scale
Load Cell Trim
Power up Zero: No
Auto Locking: 5 s
<<<
Exit
>>>
8. Select
Calibrate Scale
from the Settings Page 5 screen.
9. The display will change to show the “Number of Points” screen.
CALIBRATE SCALE
2
Number of points
at which the
scale will be
calibrated.
1 = SmartCal
1
2
3
4
5
6
7
8
9
SAVE CANCEL
0
.
DEL
10. With the “Number of Points” screen displayed, select 1, 2, 3, or 4 on the touchscreen
and then press the
Save
key.
1 = SmartCal Calibration
(calibrated load placed over each load cell only once)
2 = Two Calibration Points
(empty scale and full test load)
3 = Three Calibration Points
(empty scale, half test load, and full test load)
4= Four Calibration Points
(empty scale, one fourth test load, half test load, and full test load)
11. Proceed to the next sections of the manual for the type of calibration selected.