3300-0298-0M Rev E
y
Solace Dialysis Scale Owner’s
31
2 = Two Calibration Points
1. With the “Number of Points” screen displayed, press the 2 key on the touchscreen and
then press the
Save
key.
CALIBRATE SCALE
2
Number of points
at which the
scale will be
calibrated.
1 = SmartCal
1
2
3
4
5
6
7
8
9
SAVE CANCEL
0
.
DEL
2. The display will change to show “To calibrate zero, remove any weight from the scale
before proceeding”. Make certain the scale is empty and then press the
Save
key.
CALIBRATE SCALE
0
To calibrate zero,
remove any weight
from the scale
before proceeding
1
2
3
4
5
6
7
8
9
SAVE CANCEL
0
.
DEL
3. The display will change to show “Working” for a few seconds.
CALIBRATE SCALE
Working
To calibrate zero,
remove any weight
from the scale
before proceeding
1
2
3
4
5
6
7
8
9
SAVE CANCEL
0
.
DEL
4. The display will then change to show the “Known test weight” screen.