ENG
17
1.5
Specific mounting instructions
OPTICLUX 10-1 C T1 / OPTICLUX 10-2 C T1
Mounting material is not included.
The ceiling mounting must only be fixed to ceilings
that have a concrete stability class B25 (C20/25) or
higher.
Solid ceiling reinforcement parts must not come into
contact with the ceiling mounting. If there is any
doubt, an authorized professional must confirm that
mounting is possible on the mounting base being
used.
Drilling must be done by specialists in compliance
with the drilling tolerances for the reinforcement
anchor
that
have
been
approved
by
the
manufacturer.
Screws must be tightened carefully using a torque
key that complies with the instructions of the
reinforcement anchor manufacturer.
OPTICLUX 10-1 P TX / OPTICLUX 10-2 P TX
Mounting material is not included.
When using the accessory "wall mount", the
installation must be undertaken by a professional.
The wall must guarantee a firm hold.
Use only mounting material which is suitable for the
corresponding substrate.
2
SCOPE OF SUPPLY
2.1
OPTICLUX 10-1 C T1 / OPTICLUX 10-2 C T1
A:
1x
ceiling bracket (inner surface of B)
B:
1x
ceiling cover
C:
1x
end ring
D:
1x
ceiling tube incl. cable
E:
1x
light with adaptor (pre-assembled)
F:
1x
loupe cap
2.2
OPTICLUX 10-1 P TX / OPTICLUX 10-2 P TX
1x
light
1x
power cable
1x
loupe cap
3
ASSEMBLY of OPTICLUX 10-1 C T1 /
OPTICLUX 10-2 C T1
3.1
Load data
Bending moment M
B
25Nm
Vertical weight F
G
90N
3.2
Attaching the ceiling bracket
DANGER
Assembly by qualified personnel
Assembly must be carried out by qualified
personnel only. Lack of appropriate knowledge
could be life-threatening.
DANGER
Life-threatening danger from falling light.
The ceiling must be made of solid concrete to
guarantee a secure hold.
DANGER
Electric shocks are life threatening
It must be possible to separate the light from the
mains by an external all-pole switch (not included).
DANGER
Electric shocks are life-threatening
Voltage supply from the power connection must be
protected on all poles by means of an overcurrent
release
(according
to
national
installation
requirements; it is not included in the scope of
A
B
C
D
E
F