84
66 15 574 D3472
D3472
.
201.20.01
.
02
03.2019
5 Operation
Sirona Dental A/S
5.2 Using the menu and display information
Operating Instructions DAC Universal
The lower the sequence number, the newer the error is.
An error marked with the sequence number 1, is the newest error. While
an error marked with sequence number 18 is the oldest. The oldest error
will then be deleted from the errorbuffer whenever a new error is
detected.
5.2.2.3.5
Info
Enter this section to view the serial number and software version of the
DAC Universal.
5.2.3
Default settings
The default settings are predetermined by the manufacturer:
Restoring the default settings
To restore the default settings, please proceed as follows:
✔
The display shows the ready screen or the heating phase modus.
➢ Press the "Down"-Key and the "Left"-Key simultaneously for more
than two seconds.
The settings are restored to default, see the default settings table
above.
Program
134 °C / 273 °F, Sterilization time 3 min.
Printer:
OFF
Language:
UK
Oil expel:
1
Oil settings:
1
Automatic water:
OFF
Eco mode
ON
Brightness:
30 %