7 Errors / Display messages
Operating Instructions DAC Universal
7.1 Display messages and what they mean
båÖäáëÜ
95
The DAC Universal failed
to fill the water tank within
4 minutes.
Important: This error can occur only if the DAC
Universal is connected to a permanent water supply.
Make sure the DAC Universal is supplied with water.
If error 95 was caused by a leak, there will be a large
pool of water on the counter under the DAC
Universal.
Check for water under the DAC Universal. If this is the
case, interrupt the water supply to the DAC Universal
and please contact your authorized dealer.
96
Insufficient pressure
during expel.
Please contact your authorized dealer.
97
No water during external
cleaning.
Press the "Up" and "Right"-Keys simultaneously to
reset and then restart the DAC Universal.
If the water level is too low, fill demineralized or
distilled water (water conductivity < 3.0 µS/cm) into
the water tank.
Pump water into the chamber until water appears at
the bottom of the chamber. (See "Checking water
supply to the chamber [ → 105]"). If this is the case,
see "Drain the water tank [ → 108]".
If no water appears in the chamber, please contact
your authorized dealer.
Empty the drain tank if it is full.
Clean or change the drain filter.
Make sure the drain hose is not blocked, defective, or
kinked.
IMPORTANT! The drain hose must be made of a
heat-resistant material such as PTFE.
If the error occurs repeatedly, please contact your
authorized dealer.
Code
no.
Pictogram
Description
Corrective action