5 Operation
Operating Instructions DAC Universal
5.3 Reprocessing with the DAC Universal
båÖäáëÜ
16. Check whether the dark status bar of the chemical indicator is in the
"ACCEPT" range.
If everything is correct, reprocessing was successful.
17. Document the cycle according to the work instructions applicable in
your practice.
18. Remove all the instruments from the Flex lid when the lid and the
instruments have cooled down sufficiently. See the separate
operating instructions for adapters for detailed instructions on
removing the individual adapters.
The instruments can now be used again.
High-level disinfection of unwrapped instruments is not sufficient for
invasive procedures. Instruments must be wrapped and a sterilization
must be conducted in a class B sterilizer (e.g. DAC Premium).