6 Cleaning and maintenance
Operating Instructions DAC Universal
6.6 Cleaning the water tank
båÖäáëÜ
Water is pumped from the water tank through the chamber and into
the drain tank / drainage system.
The lid opens again automatically when the water tank is empty.
Depending on the amount of water in the water tank, this will take
approx. 10-40 min.
The display still shows the "Drain the water tank" option.
Switch off the DAC Universal to prepare for the "Cleaning the water
tank" step as described below.
6.6
Cleaning the water tank
Frequency: weekly or every 50 cycles (whichever comes first).
✔
The water tank is empty, see "Drain the water tank [ → 108]".
✔
The DAC Universal is switched off and cooled down.
1. Remove the sieve from the water tank
2. Grasp the recesses of the water tank with both hands.
3. Wiggle the water tank while gently pulling it up until it comes loose.
IMPORTANT
If you have installed the DAC Universal with a permanent water
connection, the automatic water function must be turned on again when
you have started the DAC Universal, see "Removing and cleaning the
drain tank [ → 113]".