Connecting the drain tank
If the DAC Universal is not connected directly to the drainage system, a
drain tank with a condensation tube must be installed to collect waste
water from the reprocessing process.
For connecting the drain tank, a size 6/4 mm PTFE tube must be used.
The length of the PTFE tube must not exceed 3 m. A suitable tube is
included with the drain tank.
✔
You are wearing your protective gear.
✔
The PTFE tube is at hand.
1. Make sure that the drain system is clear and no counter pressure can
build up.
2. Empty and clean the drain tank once a week or when it is full (content
reaches the highest mark on the label).
IMPORTANT
Dispose of the waste water according to the national guidelines and the
applicable regulations for waste disposal.
IMPORTANT
A drain tank other than the original one must not be used.