7 Errors / Display messages
Operating Instructions DAC Universal
7.1 Display messages and what they mean
båÖäáëÜ
73
The sterilization
temperature was not
maintained during the
back-flush process.
If the water level is too low, fill demineralized or
distilled water (water conductivity < 3.0 µS/cm) into
the water tank.
Press the "Up" and "Right"-Keys simultaneously to
reset and then restart the DAC Universal.
If the error occurs repeatedly, please contact your
authorized dealer.
74
When heating for
sterilization, a
temperature of 120 °C /
248 °F was not reached
within 2.5 minutes.
If the water level is too low, fill demineralized or
distilled water (water conductivity < 3.0 µS/cm) into
the water tank.
If the DAC Universal is in a room with a temperature
< 16 °C / 61 °F, ensure that operating conditions are
fulfilled.
Press the "Up" and "Right"-Keys simultaneously to
reset and then restart the DAC Universal.
If the error occurs repeatedly, please contact your
authorized dealer.
75
Heating for sterilization
from 120 °C / 248 °F to
134 °C / 273 °F exceeded
2.5 minutes.
If the water level is too low, fill demineralized or
distilled water (water conductivity < 3.0 µS/cm) into
the water tank.
If the DAC Universal is in a room with a temperature
< 16 °C / 61 °F, ensure that operating conditions are
fulfilled.
Press the "Up" and "Right"-Keys simultaneously to
reset and then restart the DAC Universal.
If the error occurs repeatedly, please contact your
authorized dealer.
76
Valve error
Please contact your authorized dealer.
77
Valve error
Please contact your authorized dealer.
Code
no.
Pictogram
Description
Corrective action