3 Product description
Operating Instructions DAC Universal
3.1 Functions
båÖäáëÜ
See "NitraClean cleaning tablet [ → 140]" for more information on
NitraClean cleaning tablets.
Sterilization
After external cleaning, the chamber is heated to 134 °C (273 °F) and the
instruments are sterilized. Air is extracted from the chamber and the
chamber is filled with pressurized, saturated steam.
2 different sterilization programs are available:
● 134 °C (273 °F) with a sterilization time of 3 minutes
● 134 °C (273 °F) with a sterilization time of 18 minutes
To select the program, see "Programs and program selection [ → 27]".
See "Reprocessing with the Basket lid [ → 94]" for instructions on
conditioning solid instruments.
3.1.3
Functions with the Flex lid
The unwrapped Flex lid instruments are reprocessed as follows using the
appropriate adapters on the Flex lid.