32
Service and maintenance
SWT4 User Manual v3.1
5. Now briefly open and close the valve at the syringe a few times:
water is drawn into the Tensiometer while at the same time the air
bubble is sucked into the vacuum bottle. Repeat this 2 or 3 times
until no more bubbles come out.
6. Close the valve of the vacuum bottle and remove the bottle. Inject
5 ml of water from the syringe into the refilling tube. Remove the
syringe. Put a drop of water on the end of both the rubber and
steel tubes and connect them together.