30
Service and maintenance
SWT4 User Manual v3.1
Procedure (
see fig. 4.2 & 4.3):
1. Connect the
SWT4R
to the measuring device and keep an eye on
the pressure signal all the time.
2. Two steel capillary tubes come out from the
SWT4R
shaft: the
refilling tube and the de-airing tube. In a downwards tilting
installations the marked tube is the refilling tube, in an upwards
installation the unmarked tube is the refilling tube. Pull off the
rubber tube from the refilling tube.
3. Degas the water inside the syringe as described above.
4. Connect the syringe to the refilling tube.
5. Carefully inject water into the refilling tube until no bubbles come
out of the de-airing tube. Inject at least 25 ml. Check the pressure
all the time!
6. Remove the syringe. Put a drop of water on the ends of both the
rubber and steel tubes
– and then connect them together.
4.1.4 Refilling with a vacuum pump (SWT4R only)
To achieve the maximum possible measuring range Tensiometers can
be completely degassed using a vacuum pump. This method can be
applied for installed Tensiometers at any installation angle as well as
for Tensiometers not yet installed. For refilling tubes longer than 5
meter this method should always be used.
The refill kit
SWT4-FRK2
includes all required tools: hand-operated
vacuum pump, vacuum bottle with tube and syringe with valve.
Fig. 4.4 External refill kit SWT4-FRK2 for the SWT4R