SWT4 User Manual v3.1
Service and maintenance
31
Fig. 4.5 Downwards installation
– the marked tube is the
refilling tube, the unmarked tube the de-airing tube
Fig. 4.6 Upwards installation
– the marked tube is the
de-airing tube, the unmarked tube the refilling tube
Procedure
(see Fig. 5.4 & 5.5):
1. Connect the
SWT4R
to the measuring device and keep an eye on
the pressure signal all the time.
2. In a downwards tilting installation the marked tube is the refilling
tube. In an upwards installation the unmarked tube is the refilling
tube. Pull off the rubber tube from the refilling tube.
3. Degas the water inside the syringe as described above. Connect
the syringe to the refilling tube and close the valve!
4. Connect vacuum bottle and de-airing tube. Using the vacuum
pump, evacuate the bottle to the maximum possible vacuum. This
will enlarge the remaining bubble inside the cup.