HD50
- 42 -
V1.4
Characteristics of terminal header inputs (only HD50...H):
Pt100 / Pt1000
Measuring range
-200…+650 °C
Resolution 0.1
°C
Accuracy
± 0.1 °C (excluding probe error)
Sensor coefficient
α
=0.00385 °C
-1
Connection
2, 3 or 4 wires
Thermocouple
Thermocouple type
K, J, T, N, E.
The inputs are not isolated,
use thermocouples with
isolated hot junction
.
Measuring range
type K: -200…+1370 °C
type J: -100…+750 °C
type T: -200…+400 °C
type N: -200…+1300 °C
type E: -200…+750 °C
Resolution 0.1
°C
Accuracy
(excluding probe error)
type K: ± 0.1 °C (< 600 °C)
type J: ± 0.1 °C
± 0.2 °C (> 600 °C)
type T: ± 0.1 °C
type N: ± 0.1 °C (< 600 °C)
± 0.2 °C (> 600 °C)
type E: ± 0.1 °C (< 300 °C)
± 0.2 °C (> 300 °C)
Input 0/4…20 mA
Shunt resistance
Internal (50
Ω
)
Resolution 16
bit
Accuracy
± 2 µA
Inputs -50…50 mV, 0…50 mV, 0…1 V and 0…10 V
Input resistance
100 M
Ω
Resolution 16
bit
Accuracy
± 0.01% f.s.
Potentiometric input
Potentiometer
Typically 10 k
Ω
.
Resolution 16
bit
Accuracy
± 0.01% f.s.