DeguDent SOLARIS Service Manual
11.5 Heater problems
11.5.1 No heating
Temperature of the galvanizing bath remains
unchanged
Possible causes:
Heater defective
Replace heater
11.5.2 Insufficient heating
Heating takes too long
Possible causes:
EEPROM parameter error:
-
Heating rate
too low
-
Wrong
temperature offset
or
temperature
factor
-
Check the
heating rate
EEPROM value using the
rhd
command; adjust that value using the
shd
command if required; see Sections
7.1
and
9.2.3
-
Check the
temperature offset/temperature factor
EEPROM values using the
ert
command;
adjust that value using the
ewto/ewtf
commands if required; see Sections
7.1
and
9.2.3
11.5.3 Temperature too low
Galvanization temperature significantly too low
during the process
(STATUS: Error code 35)
Possible causes:
-
Too much liquid was used
-
Heater defective
-
Heater controller defective
Use the suggested liquid volume (set in user menu)
Replace heater/heater controller as appropriate
11.5.4 Temperature too high
Galvanization temperature significantly too high
during the process
(STATUS: Error code 34)
Possible causes:
Heater controller defective
Replace the heater controller