DeguDent SOLARIS Service Manual
11.1 Power-up self-test
11.1.1 EEPROM errors
Error screen:
"** EEPROM **
EEPROM read error.
Call service technician!"
(STATUS: Error code 51)
Possible causes:
Device number in EEPROM not legible
-
Enter the serial number in EEPROM using the
ewi
command, see Section
9.2.3
-
Restart the unit
Error screen:
"** EEPROM **
EEPROM read error.
Call service technician!"
(STATUS: Error code 54)
Possible causes:
EEPROM data integrity compromised,
invalid checksum
Send in unit for service
11.1.2 Temperature sensor defective
STATUS: Error code 33
Possible causes:
Hardware problem
Replacing the temperature sensor
11.1.3 Interrupted processes/recipes
Error screen:
"WARNING
Incomplete sample(s) … in
recipe … after …% process time "
(STATUS: Error code 20)
Possible causes:
Power outage during the previous
galvanization process
-
Confirm error message using the menu OK button
-
Read the process recipe data using the
err
command, see Section
9.2.3
-
Analyze the process data, see Section 7.2.6
-
Repeat the galvanization process