DeguDent SOLARIS
Name:
Version:
Date:
Page:
Service Manual
1.8
21 Jan 2005
15 of 43
DeguDent SOLARIS Service Manual
7.2.6
Process data
The process data of the last three recipes are stored in EEPROM. A recipe counter that is also
present in EEPROM defines the number of the last recipe applied, facilitating analysis of the relevant
data for errors and process failures due to power outages. The following process data are stored for
each object:
Name
Meaning
Comment
EnblFlg
Indicates whether object is in
use
TRUE (1)
-
Object in use
FALSE (0)
-
Object not in use
RunFlg
Indicates whether process is
active
TRUE (1)
-
Process started (active)
FALSE (0)
-
Process terminated
(inactive)
ErrVal
Type of failure
Error during galvanization, see Section
11
SplType
Object type (sample type)
Subdivision into types 1…8, user selection
1
CurStart
Start current [mA]
CurEnd
End current [mA]
Characteristics of the current curve during
galvanization
Surface
Surface [mm²]
Object surface including contact surface
1
LayerThick
Layer thickness [µm]
Object-specific, user selection
GoldMass
Gold mass [mg]
Calculated total mass, object-specific
Location
Slot
Logical slot number
2
Zone
Internal or external slide ring
0
–
Internal ring
1
-
External ring
Current
Current [mA]
Last current value used
GalProcTime
Total process time
Object-specific
GalProcRstTime
Remaining process time
Object-specific
TimeStamp
Time of last updated
Date/time, can be traced back one month
The sequence of object data corresponds to the physical channel numbers. Data are not accessible
via the local menu but only via the serial command interface. Analysis of the process data can only be
performed by authorized service technicians.
1
Object type
Type
Surface (mm²)
Contact surface (mm²)
1
50,0
10
2
65,0
10
3
87,5
10
4
115,0
10
5
150,0
10
6
200,0
10
7
265,0
10
2
Logical slot numbers (seen from below):
1
2
3
4
8
7
5
6