DeguDent SOLARIS
Name:
Version:
Date:
Page:
Service Manual
1.8
21 Jan 2005
18 of 43
DeguDent SOLARIS Service Manual
9
The command interface
An alternative to local operation is operation using a serial command interface. The serial interface
parameters are as follows:
Baud rate:
19,200
Data bits:
8
Parity:
None
Stop bits:
1
Flow control:
None
There are two types of commands: standard commands and service commands. Service command
are password-protected and require operator authorization.
There are commands
-
for parameter and user settings and requests
-
for device status requests
-
for evaluating recipe data
-
for requesting the error history
-
…
A detailed command description exist for each command.
The serial port can be accessed using a standard terminal program (e.g. Hyperterminal).
9.1
Output and logging
Depending on the command issued and on the currently running processes, text output may be
available at the serial port. They represent command responses, error messages, information
messages …
9.1.1
Output modes
There are four output modes that determine the amount and type of output received.
Output mode
Description
0
No output
1
Output of galvanization parameters during process
2
Comprehensive troubleshooting output
4
Minimal output, for internal use only