DeguDent SOLARIS
Name:
Version:
Date:
Page:
Service Manual
1.8
21 Jan 2005
21 of 43
DeguDent SOLARIS Service Manual
9.2.3
Service commands
Command: SH – Set heater parameters
Command
Parameters
Explanation and examples
1…2,0…1
Set a heater: OFF/ON
Heater 1 - small
Heater 2 - large
d
Set heating rate [0.001°C/min], EEPROM entry
Example: shd,1500 -> set heating rate to 1.5°C/min
sh
g
Set galvanizing temperature [0.1°C]
Example: shg,650 -> Set galvanizing temperature to 65°C
Command: SI – Set electric current values
Command
Parameters
Explanation and examples
a,0…4095
Set all currents of all DA converters [bits]
0…7,0…4095
Set the current of one DA Converter [bits]
r,0…4095
Set the reference current [bits]
p,0…1
Set power mode: OFF/ON
0 - Power mode OFF (normal mode)
1 - Power mode ON
si
o,0…1
Set current offset: OFF/ON
0 - Offset OFF (set current without offset)
1 - Offset ON (default, set current with offset)
Command: SM – Set motor parameters
Command
Parameters
Explanation and examples
a,0…1
Setting all motors to OFF/ON
1…3,0
Setting one motor to OFF
1…3,1
Setting one motor to ON
1…3,2
Setting one motor to LEFT
sm
1…3,3
Setting one motor to RIGHT
Command: SF – Set liquid parameters
Command
Parameters
Explanation and examples
sf
0…1,<volume>
Set beaker size (0=small, 1=large) and liquid volume [ml]
Command: SQ – Set stirring level
Command
Parameters
Explanation and examples
sq
1…20
Set stirring speed [range 1..20]
Command: SK – Set display contrast
Command
Parameters
Explanation and examples
sk
1…128
Set contrast of the LCD display[range 1..128]