12
Menu Operation
Menu Operation
PosiTector
RTR-H
functions are menu controlled. Access them by
selecting
Menu
in the upper right of the PosiTector App display.
The
Main Menu
contains the following
sub-menus
...
Memory
(pg 15)
•
Statistics
(pg 23)
•
Cal Settings
(pg 13)
•
Setup
(pg 24)
•
Select
Exit
to close the menu.
Verification
Accuracy is verified using the included check shim. Ensure that the
probe has been cleaned with the included cleaning card and that the
probe has been zeroed.
Place the check shim into the measurement opening. The average of
several readings should be within the combined tolerance of both the
probe and the shim. If not, the probe may need to be returned to your
dealer for service.
The check shim is specifically intended for the PosiTector
RTR-H
. The
probe is designed to measure burnished replica tape within a limited
measuring range and automatically subtracts 50.8 um (2 mil) from
height measurements to account for the incompressible polyester film.
Therefore plastic shims intended for other instruments such as coating
thickness gages will not be measured properly.