13
Select
Zero
from the
Cal Settings
menu
1
Simultaneously press and hold both probe buttons firmly until the
2
arrows point outward. Do not place anything in the measurement
opening during this procedure.
Simultaneously press and hold both probe buttons firmly
until “0” displays (arrows do not display as in above). This can be
performed from the main measurement screen without having to
access the menu.
Reset Current Cal
Restores factory zero.
Tape Grade
Selects a replica tape grade. Required when Linearize mode (HL) is
ON.
Coarse
•
X-Coarse (default)
•
Coarse Minus and X-Coarse Plus grades are not supported
in Linearize mode since the only function of these two grades is to
improve upon the accuracy of Coarse tape at its low end and
X-Coarse tape at its high end, something the PosiTector
RTR-H
probe
will automatically do when measuring HL. When Linearize mode is
OFF, the PosiTector
RTR-H
probe will measure all four grades of
replica tape just like a conventional spring micrometer.
Shortcut:
NOTE:
Cal Settings Menu
Cal Settings Menu
Zero
The probe should be zeroed regularly to ensure best accuracy. It is
particularly useful during long measurement sessions. Clean the
probe with the included cleaning card before performing a probe
Zero
.