11
PosiTector RTR-P models
It is
important
to clean the anvils before each measurement when
obtaining peak density (P
d
) and generating surface images.
1. Ensure tape is properly positioned.
2.
Several images of the foam surface are
captured during each measurement.
Image blurring will occur if the tape is
touched or removed during this time.
Optical Grade Replica Tape
(pg. 7) provides higher quality
SDF surface data files than conventional Press-O-Film replica tape.
This special grade of tape recommended when taking
measurements which will be exported as SDF images.
NOTE:
a) Firmly press both probe buttons simultaneously and hold
until the Gage beeps and the arrows point outward.
b) Remove your fingers from the tape and release both probe
buttons (
do not
hold or remove tape from the gage).
c) Hold the gage steady while the probe measures the
compressed foam until the following graphic disappears.
3. The surface profile height (H
L
) measurement is immediately dis-
played. The tape can now be removed. An hour glass will dis-
play while peak density (P
d
) and the associated 2D/3D images are
being generated.