17
During measurement, if the linearized measurement falls outside
of the selected tape grade's range, the Gage will suggest a more
suitable tape grade to perform the measurement.
Reset
Reset
(menu reset) restores factory settings and returns the Gage
to a known condition. The following occurs:
- All batches, stored measurements, images, SDF surface data
files, batch names and screen captures are erased.
- Menu settings are returned to the following:
Date, Time and WiFi are not affected by either
Reset
.
Setup Menu
Setup Menu
NOTE:
Memory
= OFF
Statistics
= OFF
Peak Density*
= On
Memory Display
= None
Bluetooth
= OFF
WiFi
= OFF
2D*
= OFF
Tape Grade
= XC
Linearize
= ON
3D*
= OFF
*
PosiTector RTR-P models only
Units
= microns
Flip Display
= Normal
Sound
= ON
Language
= English
Battery Type
= Alkaline
Backlight
= Normal
USB Drive
= ON
Auto Sync
= OFF
Perform a more thorough
Hard Reset
as follows:
1. Power down the Gage and wait 5 seconds.
2. Simultaneously press and hold the (+) and center buttons
until the
Reset
symbol appears.
This returns the Gage to a known, “out-of-the-box” condition and
performs the same function as a menu
Reset
with the addition of:
- All calibration adjustments and Cal Memory are cleared and
returned to the Gage’s factory calibration settings.
- Bluetooth Pairing info is cleared.
- Menu settings are returned to the following: