9
Prepare the Gage
With the Gage powered-down, clean the measurement surfaces
using the included cleaning card. Swipe the card through the
opening a few times while firmly pressing both probe buttons to
remove dust particles and residual tape adhesive.
Power-up the Gage by pressing the center navigation button. An
image appears indicating that the probe requires zeroing.
Simultaneously press and hold both probe buttons firmly
until the Gage beeps and the arrows point outward. Do
not place anything in the measurement opening during
this procedure.
Measure the included check shim to verify accuracy (pg. 13).
Ensure the correct grade of replica tape, C or XC, is displayed in
the upper left corner of the display (pg. 7).
Cleaning the anvils (pg. 12) is especially important when
using the
PosiTector
RTR-P
to measure peak density (P
d
) and
generate surface images. If the cleaning card is damp, wait for the
anvils to dry before measuring.
NOTE: