Replacing the Rupture Disk Diaphragm
1.
Turn off system by toggling ON/OFF switch to OFF.
2.
From the front of the system and facing the cell a union with soft tubing will be located. (Figure
16)
3.
Slowly unthread the union halves. Inside is the diaphragm. The diaphragm is made of Teflon and
the thickness determines the rupture pressure. If it is ruptured, a hole will be seen.
4.
Remove the damaged disk and replace with a new one. Replacement disks have been furnished
with the MIOX operations kit. If more are needed contact MIOX customer service.
5.
Carefully thread the union back together with the disk centered inside.
Figure 17 Ruptured Diaphragm
Acid Washing Cell
See MIOX Acid Washing Cell Procedure (P/N: 102-00010). Contact MIOX Customer Service if copy is not