Set the Coarse Zero

 control

 

to position 8 

and 

Coarse 

Span control to 

position 0.

Position the front panel switches to the desired settings for the application. Refer to Fig. 2 
for details. Connect readout device (i.e. voltmeter) to ANO & A Common.

(b)

(c)

13

With the transducer unloaded, adjust the Coarse Zero and the Fine Zero control until the 

desired analog output is achieved. If greater zero authority is desired, place the   front panel 
Switch 7 -  "Zero Adj"  to the extended range. This will allow the user to obtain 100% zero offset control. 
The Normal postion of switch 7 allows for approx. 25% Zero authority.  

(d)

calibration resistor can be approximated

mv/v sensitivity rating of the transducer

known.

as follows, assuming that

and the bridge resistance

the

are

13

Apply a known dead weight value which is greater than one half of full scale  
in the positive direction or activate the "SHP" terminal (shunt positive) by connecting"SHP" to 
"Power Common" as described in Fig. 3. Adjust Coarse Span to obtain your nominal full scale 
analog output. Use Fine Span control to adjust the signal to the  precise value required (or for an 
analog value that correspondes with the appropriate Shunt Calibration value). Unload the 
transducer and check "Zero" and adjust as necessary.  Repeat Span  load and verify values. If the 
transducer is to be also used in the negative realm, load the transducer in the negative direction 
with the the same dead weight value or activate the "SHN"  (shunt negative) and adjust the 
Symmetry control until you obtain the proper negative output reading.

(e)

Repeat  step  (e)  as  needed  to  obtain  analog  output  precision.  Note  that  any Span (gain) 

adjustments will effect the Zero (balance) value. 

the Coarse and Fine SPAN controls. If not, adjust the Symmetry control

to obtain the negative Equivalent Input value.

If dead weight calibration is not practical and the transducer calibration

data is unknown, the Equivalent Input value for the factory-installed

calibration resistor can be approximated

mv/v sensitivity rating of the transducer

known.

as follows, assuming that

and the bridge resistance

the

are

13

(f)

(m)

When completed, ensure connections and proper shielding are correct to the moduleand re-install the top conver 
to the IL70A case.

Note: 

When appling the SHUNT resistor value for calibration, the sensor should be in an "unloaded" condition.  

3. CALIBRATION

Calibration  IL70A.3

DC Strain Gage Conditoner

 

Module

IL70A

Summary of Contents for IL70A

Page 1: ...CCW for torque cells the IL70A contains a symmetry adjustment which corrects for negative span or slope of the transducer The Model IL70A is best calibrated by means of a two point dead weight or shun...

Page 2: ...miento de este instrumento debe ser realizada por personal calificado solamente El C digo El ctrico Nacional define a una persona calificada como una que est familiarizada con la construcci n y operac...

Page 3: ...r elektrische Sicherheitsauflagen f r Angestellt Arbeitspl tze ANSI NFPA 70E 2012 der Vereinigten Staaten und alle zus tzlichen Arbeitsplatzsicherheitsauflagen folgen die auf Ihre Installation anwend...

Page 4: ...y que el material aislante no tenga rajaduras Reemplace de inmediato cualquier parte defectuosa Mesures de S curit Les mesures de s curit suivantes doivent tre prises chaque fois qu un type de connex...

Page 5: ...henden Schaltung hergestellt werden Tragen Sie Schutzbrillen und Isolierhandschuhe wenn Sie Anschl sse an den Stromkreisen vornehmen H nde Schuhe und Fu boden m ssen trocken sein wenn Sie Anschl sse a...

Page 6: ...ble up to 70mA Power Supply 11 28 Vdc regulated 2 watts max Analog Ouput selectable 0 to 5 0 to 10 Vdc 4 12 20 or 4 20mA 20 over range voltage only Operating Temperature 10 to 70 Degrees C 5 to 95 rel...

Page 7: ...Label Function 1 2 3 4 5 6 7 8 Analog Output SHP SHN NC Analog Common Power Common 24 Vdc Shield Signal Output Shunt Positive switch input Shunt Negative swith input No connection Signal Output Power...

Page 8: ...located between the two wiring connectors Below are the appropriate wiring and jumper selections depending on the application using the recommended 7 wire configuration TRANSDUCER CONNECTIONS IL70A 2...

Page 9: ...st the instrument sensitivity accordingly Place the jumper on the IL70A to INT to enable the 59K Ohm resistor A one percent 59 kilohm calibration resistor is installed in the IL70A at the factory The...

Page 10: ...o obtain your nominal full scale analog output Use Fine Span control to adjust the signal to the precise value required or for an analog value that correspondes with the appropriate Shunt Calibration...

Page 11: ...l The Remote Cal function provides a convenient method of periodically monitoring calibration of the instrument in the positive or negative realm Switch 1 Mode current I or voltage analog output Switc...

Page 12: ...by the direction of the loading or off zero reading c The inability to balance correctly where the instrument output readstotally off scale and the BALANCE controls have no authority can very likely...

Page 13: ...Fig 4 Star Bridge DC Strain Gage Conditoner Module IL70A...

Page 14: ...oduct or products for any reason the customer must call Daytronic Customer Support Services at ALL EQUIPMENT TO BE REPAIRED OR REPLACED UNDER WARRANTY MUST BE RETURNED TO THE FACTORY Once the customer...

Reviews: