background image

where

 

X

 

=

 

Equivalent Input, % of full scale

R

b

 = 

bridge resistance, ohms

K = 

transducer sensitivity, mV/V full scale

R

c

 = 

calibration resistance, ohms (59 k installed)

Sample Calculation: Assume that K = 3.000 mV/V for a 5000-pound loadcell

(fullscale) with a bridge resistance of 350 ohms.

49.4% 

of full scale = 2472 pounds

X =

25000 R

b

K R

c

X =

59000 x 3

25000 x 350

 =

If the transducer calibration is unknown, the following calculation is useful to calculate an 

Equivalent Input

 

value for the factory installed, 59K Ohm shunt resistor. This will approximate the value assuming that the 
mv/V sensititivity and bridge resistance of the sensor are known. 

3. CALIBRATION 

(cont.)

Remote Calibration Check:

 The instrument can be placed in the calibration mode

(positive or negative) by shorting pin Power Common and SHP for positive or SHN for negative  on  the 
lower  I/O  connector.  Figure  4  indicates  two  methods  of  remotely  entering  the  calibration  mode 
(external  switch,  or  push  button  control).  The  Remote  Cal  function  provides  a  convenient 
method  of  periodically  monitoring  calibration  of  the  instrument  in  the  positive  or  negative  realm.

Switch 1 - Mode - 

 current (I) or voltage analog output

Switch 2 - Volts  - 

 +/- 5 or +/- 10 Vdc when mode is voltage

Switch 3 - Current - 

 4-12-20 or 4-20mA when mode is current

Switch 4 - Filter - 

 5 kHz or 200 Hz at 3 db, for 200 Hz switch 5 must be down

 

Switch 5 - Filter - 

 10 Hz or 200 Hz at 3 db, for 200 Hz switch 4 must be down

Switch 6 - Excitation - 

 excitaton level, 2.5 or 5.0 Vdc - effects mV/V gain

 

Switch 7 - Zero Adj. - 

 extended (100%) or normal (20%) zero authority

Coarse Zero - 

16 position switch adjustment for stepped zero balance 

control with default calibration indicator.

Fine Zero - 

16 turn potentiometer for fine zero balance control

Coarse Span - 

16 position switch adjustment for stepped gain contr ol 

with default calibration indicator

Fine Span - 

16 turn potentiometer for fine gain - span control

Symmetry - 

 

adjust the negative output span to be equal to the positive output

Over Range - 

indicates the analog output is 2% greater than mode selected

 

Power - 

indicates the power input voltage is ON

Fig. 2 Front Panel Settings and Indicators

DC Strain Gage Conditoner

 

Module

IL70A

Summary of Contents for IL70A

Page 1: ...CCW for torque cells the IL70A contains a symmetry adjustment which corrects for negative span or slope of the transducer The Model IL70A is best calibrated by means of a two point dead weight or shun...

Page 2: ...miento de este instrumento debe ser realizada por personal calificado solamente El C digo El ctrico Nacional define a una persona calificada como una que est familiarizada con la construcci n y operac...

Page 3: ...r elektrische Sicherheitsauflagen f r Angestellt Arbeitspl tze ANSI NFPA 70E 2012 der Vereinigten Staaten und alle zus tzlichen Arbeitsplatzsicherheitsauflagen folgen die auf Ihre Installation anwend...

Page 4: ...y que el material aislante no tenga rajaduras Reemplace de inmediato cualquier parte defectuosa Mesures de S curit Les mesures de s curit suivantes doivent tre prises chaque fois qu un type de connex...

Page 5: ...henden Schaltung hergestellt werden Tragen Sie Schutzbrillen und Isolierhandschuhe wenn Sie Anschl sse an den Stromkreisen vornehmen H nde Schuhe und Fu boden m ssen trocken sein wenn Sie Anschl sse a...

Page 6: ...ble up to 70mA Power Supply 11 28 Vdc regulated 2 watts max Analog Ouput selectable 0 to 5 0 to 10 Vdc 4 12 20 or 4 20mA 20 over range voltage only Operating Temperature 10 to 70 Degrees C 5 to 95 rel...

Page 7: ...Label Function 1 2 3 4 5 6 7 8 Analog Output SHP SHN NC Analog Common Power Common 24 Vdc Shield Signal Output Shunt Positive switch input Shunt Negative swith input No connection Signal Output Power...

Page 8: ...located between the two wiring connectors Below are the appropriate wiring and jumper selections depending on the application using the recommended 7 wire configuration TRANSDUCER CONNECTIONS IL70A 2...

Page 9: ...st the instrument sensitivity accordingly Place the jumper on the IL70A to INT to enable the 59K Ohm resistor A one percent 59 kilohm calibration resistor is installed in the IL70A at the factory The...

Page 10: ...o obtain your nominal full scale analog output Use Fine Span control to adjust the signal to the precise value required or for an analog value that correspondes with the appropriate Shunt Calibration...

Page 11: ...l The Remote Cal function provides a convenient method of periodically monitoring calibration of the instrument in the positive or negative realm Switch 1 Mode current I or voltage analog output Switc...

Page 12: ...by the direction of the loading or off zero reading c The inability to balance correctly where the instrument output readstotally off scale and the BALANCE controls have no authority can very likely...

Page 13: ...Fig 4 Star Bridge DC Strain Gage Conditoner Module IL70A...

Page 14: ...oduct or products for any reason the customer must call Daytronic Customer Support Services at ALL EQUIPMENT TO BE REPAIRED OR REPLACED UNDER WARRANTY MUST BE RETURNED TO THE FACTORY Once the customer...

Reviews: