3. CALIBRATION

Calibration of the IL70A

 is 

accomplished by the conventional shunt technique, using an internally 

installed calibration resistor, or via the "dead-weight" method. 

Calibration Resistor.   

If a fixed resistor is shunted across one arm of a strain gage  bridge,  it  produces  an 

unbalance  equivalent  to  that  of  a  particular  value  of  mechanical input. If this Equivalent Input value is 
accurately known, it can be used as  a  reference  point  for  shunt  calibration  of  the  system.  Upon 
completion  of  installation  of  the  transducer  and  its  associated  cabling,  the  user  can:

Perform  an  overall  dead weight  calibration  using  a  precisely  known  value 

of  mechanical  input.  The  calibration  can  then  be  transferred  to  the  installed 
calibration  resistor  for  convenience  in  subsequent  checking.

Use  an  external  shunt  resistor,  please  refer  to  the  notes  on  page  one  of  this  manual.  T

he 

installed  calibration  resistor 

is  typically

  supplied  by  the  transducer  manufacturer 

or  is 

referenced  on  the  calbiraiton  sheet  that  accompanies  the  tranducer.  Using  a  known  shunt 
resistorance  can

  achieve  a  precisely  known  Equivalent  Input  allowing  the  instrument 

sensitivity  to  be  adjusted  correctly.

  Note,  this  shunt  resistor  may  reside  in  the  transducer 

itself and is accessed via the input pins of the sensor.

D

etermine  the  Equivalent  Input  value  for  the 

internally 

installed 

59K 

calibration  resistor, 

knowing the transducer sensitivity, and adjust the instrument sensitivity accordingly.

 Place the 

jumper on the IL70A to "INT" to enable the 59K Ohm resistor.

A one percent, 59-kilohm calibration resistor is installed in the IL70A at the factory. The  installed 
resistor  can  usually  be  used  even  though  the  transducer  calibration data  mentions  some 
other  resistance  value.  In  Section  4  of  this  manual,  the  techniques  described  above  are 
demonstrated.  If,  however,  the  installed  value  of calibration resistor is not appropriate 
for the transducer and 

the 

measurement range to be  used, 

an

 

external  "EXT"  shunt

  resistor 

should  be 

installed

.  The 

external 

calibration resistor  is 

installed

 

from  the  "CAL  Sense"

 

terminal 

to  either  the  +  Signal  or  -  Signal  teminals  depending  on  your  polarity  direction 

when  using  the  provided  "shunt"  switch

Note: 

A  variety  of  DC  Load  Cells  and  Slip  Ring 

Torque  Transducers are  supplied with  the  appropriate  calibration  resistor  integral  to  the 
transducer.  When  this type  of  transducer  is  used  it  will  be  necessary  to  place 

the jumper in 

"EXT" and connect the "Cal Sense" line to the appropiate pin on the transducer.

 

(1)

(2)

(3)

Calibration  IL70A.3

This  section  contains  the  instructions  for  calibrating  the  IL70A.  Included  is  a functional 
description  of  the  instrument  front-panel  (see  Figure  2).  To  perform calibration,  proceed 
as  follows  with  the  sensor  or  calibration  standard  connected.

Turn power ON to the IL70A DC Power input terminal (11 to 28 Vdc input) The front-

panel indicator should light green to indicate the application of DC  power.  Allow  10 
minutes  of  warmup  for  stabilization  of  transducer characteristics. 

(a)

the Coarse and Fine SPAN controls. If not, adjust the Symmetry control

to obtain the negative Equivalent Input value.

If dead weight calibration is not practical and the transducer calibration

data is unknown, the Equivalent Input value for the factory-installed

calibration resistor can be approximated

mv/v sensitivity rating of the transducer

known.

as follows, assuming that

and the bridge resistance

the

are

13

(m)

the Coarse and Fine SPAN controls. If not, adjust the Symmetry control

to obtain the negative Equivalent Input value.

If dead weight calibration is not practical and the transducer calibration

data is unknown, the Equivalent Input value for the factory-installed

calibration resistor can be approximated

mv/v sensitivity rating of the transducer

known.

as follows, assuming that

and the bridge resistance

the

are

13

(m)

recommended calibration resistor.

the Coarse and Fine SPAN controls. If not, adjust the Symmetry control

to obtain the negative Equivalent Input value.

If dead weight calibration is not practical and the transducer calibration

data is unknown, the Equivalent Input value for the factory-installed

calibration resistor can be approximated

mv/v sensitivity rating of the transducer

known.

as follows, assuming that

and the bridge resistance

the

are

13

(m)

Calibration 5M70.3

DC Strain Gage Conditoner

 

Module

IL70A

Summary of Contents for IL70A

Page 1: ...CCW for torque cells the IL70A contains a symmetry adjustment which corrects for negative span or slope of the transducer The Model IL70A is best calibrated by means of a two point dead weight or shun...

Page 2: ...miento de este instrumento debe ser realizada por personal calificado solamente El C digo El ctrico Nacional define a una persona calificada como una que est familiarizada con la construcci n y operac...

Page 3: ...r elektrische Sicherheitsauflagen f r Angestellt Arbeitspl tze ANSI NFPA 70E 2012 der Vereinigten Staaten und alle zus tzlichen Arbeitsplatzsicherheitsauflagen folgen die auf Ihre Installation anwend...

Page 4: ...y que el material aislante no tenga rajaduras Reemplace de inmediato cualquier parte defectuosa Mesures de S curit Les mesures de s curit suivantes doivent tre prises chaque fois qu un type de connex...

Page 5: ...henden Schaltung hergestellt werden Tragen Sie Schutzbrillen und Isolierhandschuhe wenn Sie Anschl sse an den Stromkreisen vornehmen H nde Schuhe und Fu boden m ssen trocken sein wenn Sie Anschl sse a...

Page 6: ...ble up to 70mA Power Supply 11 28 Vdc regulated 2 watts max Analog Ouput selectable 0 to 5 0 to 10 Vdc 4 12 20 or 4 20mA 20 over range voltage only Operating Temperature 10 to 70 Degrees C 5 to 95 rel...

Page 7: ...Label Function 1 2 3 4 5 6 7 8 Analog Output SHP SHN NC Analog Common Power Common 24 Vdc Shield Signal Output Shunt Positive switch input Shunt Negative swith input No connection Signal Output Power...

Page 8: ...located between the two wiring connectors Below are the appropriate wiring and jumper selections depending on the application using the recommended 7 wire configuration TRANSDUCER CONNECTIONS IL70A 2...

Page 9: ...st the instrument sensitivity accordingly Place the jumper on the IL70A to INT to enable the 59K Ohm resistor A one percent 59 kilohm calibration resistor is installed in the IL70A at the factory The...

Page 10: ...o obtain your nominal full scale analog output Use Fine Span control to adjust the signal to the precise value required or for an analog value that correspondes with the appropriate Shunt Calibration...

Page 11: ...l The Remote Cal function provides a convenient method of periodically monitoring calibration of the instrument in the positive or negative realm Switch 1 Mode current I or voltage analog output Switc...

Page 12: ...by the direction of the loading or off zero reading c The inability to balance correctly where the instrument output readstotally off scale and the BALANCE controls have no authority can very likely...

Page 13: ...Fig 4 Star Bridge DC Strain Gage Conditoner Module IL70A...

Page 14: ...oduct or products for any reason the customer must call Daytronic Customer Support Services at ALL EQUIPMENT TO BE REPAIRED OR REPLACED UNDER WARRANTY MUST BE RETURNED TO THE FACTORY Once the customer...

Reviews: