The Model IL70A I/O CONNECTIONS are via non-removable screw terminals which will accept  wire sizes from 
AWG 16 to 30.  NOTE: The recommended transducer cabling would be eight wire, individually shielded, 
twisted pair - wired as indicated (Fig. 1). Sense lines must be connected at the transducer (as recommended) 
or at the IL70A  screw terminals - as a minimum. Table 2 denotes screw terminal assignments 

TRANSDUCER CONNECTIONS  IL70A.2

DC Strain Gage Conditoner

 

Module

IL70A

Fig. 1  Model IL70A Transducer Cabling

Table 2  Model IL

70

A Pin Assignments

I/O Connector

Conditioner

Terminal

Line

Screw

Terminal

Label 

   Function

1

2

3

4

5

6

7

8

Analog Output

 

SHP

SHN

NC

Analog Common

 

Power Common

+ 24 Vdc

 

Shield

+ Signal Output

Shunt Positive

 switch input

 

Shunt Negative swith input

 

No connection

- Signal Output

Power Supply Common

 

11 - 28 Vdc Input Power

 

Earth ground

1

2

3

4

5

6

7

8

+

 Excitation

+

 Sense

Excitation

-

 Sense

+

 

Signal

- Signal

Cal Sense

Shield

+

 Sensor Power

Sensor Power Sense

-

 

Sensor Power

-

 

Sensor Power Sense

+ Signal from Sensor

- Signal from Sensor

Calibration Shunt Sense

Earth ground

J1

J2

7 Wire confirguration (

recommended

)

 

4 twisted pair, shielded wire used.
Paired as follows:

+ & - Excitation

+ & - Sense

+ & - Signal

Cal Sense and unused wire

Shield at the instrument only

4 Wire confirguration 

2 twisted pair, shielded wire used.

Paired as follows:

+ & - Excitation

+ & - Signal

Shield at the instument only

Sense lines 

must

 be connected at the 

instrument 

Summary of Contents for IL70A

Page 1: ...CCW for torque cells the IL70A contains a symmetry adjustment which corrects for negative span or slope of the transducer The Model IL70A is best calibrated by means of a two point dead weight or shun...

Page 2: ...miento de este instrumento debe ser realizada por personal calificado solamente El C digo El ctrico Nacional define a una persona calificada como una que est familiarizada con la construcci n y operac...

Page 3: ...r elektrische Sicherheitsauflagen f r Angestellt Arbeitspl tze ANSI NFPA 70E 2012 der Vereinigten Staaten und alle zus tzlichen Arbeitsplatzsicherheitsauflagen folgen die auf Ihre Installation anwend...

Page 4: ...y que el material aislante no tenga rajaduras Reemplace de inmediato cualquier parte defectuosa Mesures de S curit Les mesures de s curit suivantes doivent tre prises chaque fois qu un type de connex...

Page 5: ...henden Schaltung hergestellt werden Tragen Sie Schutzbrillen und Isolierhandschuhe wenn Sie Anschl sse an den Stromkreisen vornehmen H nde Schuhe und Fu boden m ssen trocken sein wenn Sie Anschl sse a...

Page 6: ...ble up to 70mA Power Supply 11 28 Vdc regulated 2 watts max Analog Ouput selectable 0 to 5 0 to 10 Vdc 4 12 20 or 4 20mA 20 over range voltage only Operating Temperature 10 to 70 Degrees C 5 to 95 rel...

Page 7: ...Label Function 1 2 3 4 5 6 7 8 Analog Output SHP SHN NC Analog Common Power Common 24 Vdc Shield Signal Output Shunt Positive switch input Shunt Negative swith input No connection Signal Output Power...

Page 8: ...located between the two wiring connectors Below are the appropriate wiring and jumper selections depending on the application using the recommended 7 wire configuration TRANSDUCER CONNECTIONS IL70A 2...

Page 9: ...st the instrument sensitivity accordingly Place the jumper on the IL70A to INT to enable the 59K Ohm resistor A one percent 59 kilohm calibration resistor is installed in the IL70A at the factory The...

Page 10: ...o obtain your nominal full scale analog output Use Fine Span control to adjust the signal to the precise value required or for an analog value that correspondes with the appropriate Shunt Calibration...

Page 11: ...l The Remote Cal function provides a convenient method of periodically monitoring calibration of the instrument in the positive or negative realm Switch 1 Mode current I or voltage analog output Switc...

Page 12: ...by the direction of the loading or off zero reading c The inability to balance correctly where the instrument output readstotally off scale and the BALANCE controls have no authority can very likely...

Page 13: ...Fig 4 Star Bridge DC Strain Gage Conditoner Module IL70A...

Page 14: ...oduct or products for any reason the customer must call Daytronic Customer Support Services at ALL EQUIPMENT TO BE REPAIRED OR REPLACED UNDER WARRANTY MUST BE RETURNED TO THE FACTORY Once the customer...

Reviews: