General Service Guide
69
Document No. 896 623
Press each membrane key on the module front panel for at least one second. Check that
the key being pressed is identified in the menu, i.e. text
OFF
changes to
ON
.
?
61.
Attach an adult NIBP cuff onto your arm and perform one NIBP measurement. Check the
module identifies the cuff, i.e. the text ‘Adult’ appears on the NIBP digit field for a short
time.
Check that the module gives a reasonable measured result.
?
NeuroMuscular Module, M
-
NMT
62.
Enter the
NMT
module service menu:
Monitor Setup
- Install/Service (password 16-4-34) -
Service (password 26-23-8) - Parameters - NMT
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 50 per second. Check that the module’s memories have passed the internal
memory test, i.e. ‘RAM’, ‘ROM’ and ‘EEPROM’ all state
OK
.
?
63.
Check the module membrane
Start-Up
and
Stop/Continue
keys.
Press each key for at least one second and check that the key being pressed is identified,
i.e. the color under the corresponding text in the menu turns blue.
?
Nellcor Compatible Saturation module, M-NSAT
64.
Enter the
M-NSAT
service menu:
Monitor Setup
- Install/Service
(password 16-4-34)
-
Service
(password 26-23-8)
- Parameters - M-NSAT
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 50 per second. Check that the NSAT module’s ROM memory has passed the
internal memory test, i.e. ‘ROM’ shows
OK
.
?
65.
Check that the SpO
2
probe related status information in the menu is correct. Only the
‘NoProbe’ should be active (1) when no probe is connected.
?